Agreement of total corneal power between 2 swept-source optical coherence tomography and Scheimpflug tomography in normal and keratoconic patients

To evaluate agreement of total corneal power (TCP) measured by swept-source anterior segment optical coherence tomography (CASIA 2), Scheimpflug tomography (Pentacam AXL), and swept-source optical biometer (IOLMaster 700) in normal and keratoconic patients. This is a prospective observational study conducted at King Chulalongkorn Memorial Hospital, Bangkok, Thailand. Biometric values were measured by each device for three times by two operators to evaluate repeatability and reproducibility of TCP. The agreement of TCP and other parameters including total corneal astigmatism, anterior keratometry, anterior corneal astigmatism, posterior keratometry, posterior corneal astigmatism, anterior chamber depth, white-to-white corneal diameter (WTW), central corneal thickness, and intraocular power were also evaluated. This study enrolled 100 healthy participants and 34 patients with keratoconus. The repeatability and reproducibility of TCP were high in all devices (ICC > 0.9). The agreement of TCP was excellent among three devices in both groups (ICC > 0.9). However, the agreement of TCP between CASIA 2 and IOLMaster 700 was slightly lower in healthy participants (ICC = 0.867) and k

source: PubMed Central Open Access Subsetlicence: Per-article PMC Open Access licencedatabase: /dbs/pmc.dbmode: tablerecords: 110356 (max_rowid)citation: https://www.ncbi.nlm.nih.gov/pmc/

To evaluate agreement of total corneal power (TCP) measured by swept-source anterior segment optical coherence tomography (CASIA 2), Scheimpflug tomography (Pentacam AXL), and swept-source optical biometer (IOLMaster 700) in normal and keratoconic patients.

This is a prospective observational study conducted at King Chulalongkorn Memorial Hospital, Bangkok, Thailand. Biometric values were measured by each device for three times by two operators to evaluate repeatability and reproducibility of TCP. The agreement of TCP and other parameters including total corneal astigmatism, anterior keratometry, anterior corneal astigmatism, posterior keratometry, posterior corneal astigmatism, anterior chamber depth, white-to-white corneal diameter (WTW), central corneal thickness, and intraocular power were also evaluated.

This study enrolled 100 healthy participants and 34 patients with keratoconus. The repeatability and reproducibility of TCP were high in all devices (ICC > 0.9). The agreement of TCP was excellent among three devices in both groups (ICC > 0.9). However, the agreement of TCP between CASIA 2 and IOLMaster 700 was slightly lower in healthy participants (ICC = 0.867) and keratoconic patients (ICC = 0.852) with mean differences of more than 1.0 diopter is clinically significant. Wider 95% limit of agreement were found in keratoconic eyes. Most of other parameters showed good to excellent agreement except WTW which showed poor to moderate agreement in both groups. IOL power showed clinically significant differences in patients with keratoconus.

The agreement of TCP measured by three devices was excellent in normal and keratoconic patients. However, TCP cannot be used interchangeably between devices.

Read full text (20,129 characters)

Fields

journalPLoS ONE
year2022
pmid35609043
doi10.1371/journal.pone.0268856
licenseThis is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. https://creativecommons.org/licenses/by/4.0/
source_urlhttps://pmc-oa-opendata.s3.amazonaws.com/PMC9129016.1/PMC9129016.1.xml
article_typeresearch-article
source: PubMed Central Open Access Subsetlicence: This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. https://creativecommons.org/licenses/by/4.0/year: 2022id: PMC9129016citation: https://pmc-oa-opendata.s3.amazonaws.com/PMC9129016.1/PMC9129016.1.xml

JSON · full text · full-text JSON

Related services and records

Services

  • Estate — measured countries, lanes and facts
  • Intel — live corroborated events
  • Legal — statutes, cases and citations
  • Encyclopedia — articles with inline sources
  • Space — launches, satellites and orbits
  • Search — federated search and reports
  • Pass — passport account and SSO
  • Corpora — read-only corpus origins
  • e01 — self-hosted corpora and tools

Text and metadata only; each target carries its own sources. Missing data is published as no-data, never estimated.