[Thoracic radiography in intensive care units. Comparison of an asymmetric film-screen-grid combination with an L-film].

PMID 7670010Journal RoFo : Fortschritte auf dem Gebiete der Rontgenstrahlen und der NuklearmedizinYear 1995Authors 4DOI 10.1055/s-2007-1015955

  • Source PubMed
  • Journal RoFo : Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin
  • Year 1995
  • Authors G Luska; H von Boetticher; B Amling; R Herzog
  • Licence NLM/PubMed terms

Abstract

Using an asymmetrical screen-film-grid combination with an ITC-VHC special grid (Kodak), the effect of the individual components was evaluated with regard to detail in the lungs and mediastinum as well as patient dose. The system was compared with the Curix Ortho screen film system (Agfa L-film) using phantoms and 62 patients in an intensive care unit. The results were analysed by three independent observers. Dose measurements on phantoms were carried out with a WD 10 (Wellhoefer) dosimeter and with a thermoluminescence dosimeter for patients.

Both phantom and patient images showed improved demonstration of mediastinal structures with similar details of lung structure. Details in the peripheral lungs are significantly better shown with the ITC-VHC system with the special grid than they are without the grid. However, there is an increase about three to four times in radiation dose when using the grid.

Our trial shows that the use of the ITC-VHC grid system is to be recommended for routine clinical use because of the better demonstration of details, but one must be aware of the increased radiation dose to the patient.

Record details

PMID7670010
PMCIDnot held
DOI10.1055/s-2007-1015955
JournalRoFo : Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin
Year1995
AuthorsG Luska; H von Boetticher; B Amling; R Herzog
Licencenot stated
MeSHEvaluation Studies as Topic; Humans; Intensive Care Units; Lung; Mediastinum; Models, Structural; Radiation Dosage; Radiographic Image Enhancement; Radiography, Thoracic; X-Ray Film; X-Ray Intensifying Screens