pubmed· e01.org

PubMed · PMID 7652004

A new method to determine ratios of electron stopping powers to an improved accuracy.

PMID 7652004Journal Physics in medicine and biologyYear 1995Authors 3DOI 10.1088/0031-9155/40/5/002

  • Source PubMed
  • Journal Physics in medicine and biology
  • Year 1995
  • Authors D T Burns; S Duane; M R McEwen
  • Licence NLM/PubMed terms

Abstract

A new method is presented to determine the ratio of electron stopping powers which is effective in the transfer of absorbed dose from one medium to another. The method involves an accurate measurement of the electron range in each of the media combined with a full Monte Carlo simulation of each experimental geometry. For the specific case of graphite and water, the uncertainty attainable is estimated to be around +/- 0.5% at the 95% confidence level, which is approximately a factor of three better than the best methods currently in use.

Record details

PMID7652004
PMCIDnot held
DOI10.1088/0031-9155/40/5/002
JournalPhysics in medicine and biology
Year1995
AuthorsD T Burns; S Duane; M R McEwen
Licencenot stated
MeSHBiophysical Phenomena; Biophysics; Computer Simulation; Electrons; Graphite; Humans; Methods; Models, Biological; Monte Carlo Method; Radiometry; Radiotherapy Dosage; Water

Source record

Open the PubMed record. This page holds the abstract and citation fields for this PMID; no PMC full text is held for it.

Source
PubMed / NLM
PMID
7652004
Rendered
2026-11-22
Completeness
abstract and citation fields